Specifications

2300-482
Verification
Software Overview
The 2300-482 verification software package is included with the 366X
verification kit and it is also available separately. Complete setup and op
-
erating instructions are packaged with the software in a software user
guide in Acrobat pdf format.
This software supports all 2, 3, and 4 Port VNMS instruments (A, B, and
D models) operating up to 9 GHz. The model and serial number of all of
the equipment used will be collected and recorded on electronic test re
-
cords. A 12-term (or higher) calibration will be performed under program
control.
NOTE
Special GPIB settings are required for software control.
Refer to the software user guide for complete setup
information.
All devices from the verification kit will be measured under local lockout
using graph types of Log Magnitude, Phase, Linear Magnitude, and Real.
All possible S-parameters for the given VNMS are tested with each de-
vice.
Data received from the VNMS for each device will be compared to the
characterization data stored on the floppy diskette for your kit. Each
measurement point will be checked for a value that is within the com-
puted allowable uncertainty range. Any data point that falls outside of
the allowable range will be clearly designated as a failed data point.
All measurement data, standard device characterization data, and allow-
able uncertainties are viewable after the device is measured and the data
is saved as a text file. This text file may be printed from within the pro
-
gram environment and is saved on the hard drive of the PC.
For more details on this software or the verification process, refer to the
software user guide that is packaged with the software.
4-6 MS462XX MM
FOR THE MS462XA/B/D MODELS PERFORMANCE VERIFICATION