Specifications
3-13 HARMONIC
OUTPUT CHECK
This procedure checks the harmonic output.
Equipment
Required:
q
Anritsu Spectrum Analyzer Model MS2602A
q
Anritsu 68x47B or 69x47A Synthesized Generator
q
Anritsu ML243XA Power Meter and Associated Power Sensor
Procedure: Set up the Synthesized Generator to output a CW RF signal at the fre
-
quency shown below:
At CW Frequency
Generator Output Level
600 MHz
Phase Harmonic
Standard 2nd and 3rd
Harmonic Outputs
2000 MHz
Phase Harmonic
Standard 2nd and 3rd
Harmonic Output
s
0 dBm -20 dBc £ x £-50 dBc -20 dBc £ x £-50 dBc
+ 5 dBm -20 dBc
£ x £-45 dBc -20 dBc £ x £-45 dBc
+10dBm
-20 dBc
£ x £-45 dBc -20 dBc £ x £-45 dBc
PERFORMANCE TESTS HARMONIC OUTPUT CHECK
MS462XX MM 3-39/3-40