Specifications
4. Follow the selection menu and use the corresponding soft key to
make the appropriate selections as listed below:
PERFORM CAL: 2 PORT
CAL METHOD: STANDARD
LINE TYPE: COAXIAL
SELECT CALIBRATION TYPE: 12 TERM
USE OF ISOLATION IN CALIBRATION: EXCLUDE
FREQUENCY RANGE OF CALIBRATION: 50 MHz to 3000 MHz
(50 MHz to 6000 MHz for Option 4B)
SELECT CALIBRATION DATA POINTS: NORMAL
DATA POINTS: 51 POINTS
PORT 1 CONNECTOR TYPE: SMA (M)
PORT 2 CONNECTOR TYPE: SMA (M)
LOAD TYPE: BROADBAND LOAD
TEST SIGNAL/PORT 1 POWER: -15 dBm
5. After the selections are complete, press the START CAL soft key to
begin the calibration.
6. Install the calibration device per the instructions on the display. Se-
lect the appropriate soft key to measure the calibration device.
7. When the “CALIBRATION SEQUENCE COMPLETED” message is
displayed, press the Enter key to continue.
8. Press the Display key, then select:
DISPLAY MODE
SINGLE CHANNEL
9. Press the Ch 1 key, select DISPLAY/GRAPH TYPE, and change the
graph type to LOG MAGNITUDE display.
10. Press the Meas key, select S-PARAMETER/S21.
11. Install a ZJL-4G or ZJL-6G Amplifier between Port 1 and Port 2.
12. Press the Display key and allow two sweeps to complete.
13. Select:
TRACE MEMORY
STORE DATA TO MEMORY
DISK OPERATIONS
SAVE MEMORY TO HARD DISK
CREATE NEW FILE
NOISE FIGURE MEASUREMENT CONFIDENCE CHECK PERFORMANCE TESTS
3-32 MS462XX MM
NOTE
When measuring the
throughline, replace
the 33LLF50
insertable with a
33LFLF50 insertable.
After the throughline is
measured, replace the
33LFLF50 insertable
with the 33LLF50
insertable.