Specifications

6. Select:
FROM HARD DISK/INTERNAL ENR EXTENSION TABLE
Test Procedure: Press the Appl key.
1. Select:
CHANGE APPLICATION SETUP
MEASUREMENT TYPE/TRANSMISSION AND REFLECTION
2. Install the cables and adapters as shown in Figure 3-11.
3. Press the Cal key to begin measurement calibration.
PERFORMANCE TESTS NOISE FIGURE MEASUREMENT CONFIDENCE CHECK
MS462XX MM 3-31
Use adapters
to convert
these test
ports to
3.5 mm male
connectors
Install a 33LLF50
phase equal insertibe
to port 2
Install this 33LFLF50 in
place of the 33LLF50
during the through
calibration step. Re-install
the 33LLF50 after the
calibration is complete.
Port 1
Port 2
Figure 3-11. Noise Figure Calibration Setup
CAUTION
It is very important to load these ENR table files into the instrument. The data
included in these files is used by the instrument to perform noise Figure
measurements.
NOTE
If the internal ENR
extension table file is
stored on a floppy
disk, select FROM
FLOPPY
DISK/INTERNAL
ENR EXTENSION
TABLE instead. Refer
to Section 3-7 of this
manual.