Specifications

5. Press the Config key.
6. Select:
DATA POINTS/51 MAX PTS
7. Connect the 15 dB ENR Noise Source to Port 2.
8. Verify that the measured Noise Figure is within the range specified
in Table 3-8.
9. Press the Display key.
10. Select:
TRACE MEMORY
STORE DATA TO MEMORY
VIEW DATA AND MEMORY
11. Press the Appl key.
12. Select:
NOISE Figure SETUP
NOISE SOURCE/INTERNAL
13. Connect the noise source to the Noise In connector on the rear panel.
NOISE FIGURE MEASUREMENT CAPABILITY (OPTION 4) PERFORMANCE TESTS
3-28 MS462XX MM
Model Lower Limit (dB) Upper Limit (dB)
MS4622B
50 MHz to 3 GHz 1.75 7.00
MS4623B
50 MHz to 3 GHz
3 GHz to 6 GHz
1.75
1.75
7.00
8.75
MS4624B
50 MHz to 3 GHz
3 GHz to 6 GHz
1.75
1.75
7.00
10.75
All D Models
50 MHz to 3 GHz
3 GHz to 6 GHz
1.75
1.75
11.00
14.00
Table 3-8. Noise Figure Specifications