Chapter 6 FAQ
No.
Failure
Phenomenon
Cause Analysis
Processing
1
The screen shows a
black screen
Screen motherboard
damage
Must replace the screen
motherboard, please contact with
the supplier or manufacturer
If the screen lock
function is
incorrectly
operated, the
screen enters the
black screen
standby mode
Close the lock screen, and enter
the test interface directly after
the startup
Screen damage
Need to replace the screen, please
contact the supplier or
manufacturer
Software prompts
"Abnormal
auxiliary heating
Auxiliary
temperature
selftest abnormal
Please restart the instrument
first to confirm, if still can not
solve please contact the supplier
or manufacturer
Abnormal heat-up
curve of hot cover
Power supply
problems
Verify that the power is plugged in
properly
Hot cover assembly
problems
Please restart the instrument
first to confirm, if still can not
solve please contact the supplier
or manufacturer
2
Unable to turn on
Power supply
problems
Verify that the power is plugged in
properly
Switch or power
cable damage
Please restart the instrument
first to confirm, if still can not
solve please contact the supplier
or manufacturer
3
USB flash drive
export failed
The USB disk is not
in good contact
Reinsert the USB flash drive for
confirmation
Can try to use WIFI, Bluetooth
export function
If the problem persists, contact
the supplier or manufacturer
4
HID connection
exception
Screen
communication
exception
Please restart the instrument
first to confirm, if still can not
solve please contact the supplier
or manufacturer
5
No experimental
Incorrect setting
Verify that the thermal cycling