User guide

Page 81
14
Warning: Basal Delivery Suspended
Cause Basal Edit was not saved.
Effect Basal delivery stopped.
Message Displayed when manually awakened until confirmed.
Action
Press OK to select Edit Basal. Review basal edits and
select Save/Review.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Suspend
Cause Pump suspended manually.
Effect All deliveries stop.
Message
Displayed each time pump is awakened until confirmed
and once every 15 minutes until action is taken.
Action Press OK to confirm. Resume delivery.
Beeps/Vib
User selected, once every 15 min. No progression if
confirmed each time displayed. Sweep/vibe within
one hour if not confirmed.
Warning: No Cartridge Detected, Deliveries Disabled
Cause
No cartridge detected after “Load cartridge” step
during Rewind/Prime.
Effect No deliveries.
Message
Displayed until confirmed or until pump goes to sleep.
Action
Press OK to confirm. Be sure Rewind/Prime sequence
is completed with cartridge properly in place.
Beeps/Vib
User selected, once every 3 min. No progression if
confirmed each time displayed. Sweep/vibe within
one hour if not confirmed.
Alerts, Warnings and Alarms