User guide

Page 61
11
7. To move to the next BG Target screen, scroll to the forward arrows and press OK.
8. Scroll up to the time field and press OK to change to flashing cursor for Edit mode.
9. Use the Up/Down arrows to change the segment start time. Press OK.
10. Scroll to the BG Target field. Press OK to change to flashing cursor for Edit mode.
11. Use the Up/Down arrows to change to desired setting.
12. Press OK when setting is made.
13. Scroll to the range field. Press OK to change to flashing cursor for Edit mode.
14. Use the Up/Down arrows to change the range as desired. Press OK. Repeat to set
remaining segments per your healthcare team’s recommendations.
To review your settings, highlight the forward arrows and press OK to scroll through
each segment. Confirm the times and setting values are correct.
When finished, scroll to Done and press OK to return to the Setup Adv screen.
To return to the Home screen, scroll to Home and press OK.
Setup Advanced 8 – Insulin on Board Setting
Even with rapid-acting insulin, your body takes some time to use your entire bolus.
This feature allows you to program the rate at which your body uses your bolus.
When this feature is activated and you give a bolus, your pump will tell you how much Insulin
on Board (IOB) is currently remaining and will calculate a decreased bolus dose as an option.
This helps to prevent “stacking” insulin and can help reduce your risk of hypoglycemia*.
The duration of insulin action varies from person to person and can vary based on the infusion
site you have selected and your activity level, among other factors. Your healthcare team will
give you a recommended duration to program into your Animas® 2020.
This feature is intended for use only with U100 rapid-acting insulin analogs such
as Novolog®, Humalog® or Apidra®. If you use an insulin other than Novolog,
Humalog or Apidra, do not use this feature. Use of any insulin with lesser or
greater concentration can result in serious injury or death.
This screen allows you to:
Turn the IOB feature on or off•
Select the duration•
1. Use the Up/Down arrows to scroll to the desired field.
2. Press OK to change to flashing cursor for Edit mode.
3. Use the Up/Down arrows to change to desired setting.
4. Press OK when setting is made.
To move to the next Setup Advanced screen, scroll to the bottom of the screen and
highlight Next. Press OK.