Instruction manual
ISP-500C
Fit air-muffler
~
_ / Remove plug
Air-ilush
port
4.2.2 Start-up and
shut-down
CD
Start vacuum pump according to 4.1.1 Operation [page 21].
® Stop vacuum pump according to 4.1.2 Shut-down[page 21].
Important
Continuous
evacuating
of
humid
gas
When evacuating vacuum chamber while humidity
in
chamber
is high, moisture volume drawn into pump
differs according to temperatu
re
and pressure in chamber.
When pumping vacuum chamber containing humid gas, be sure to open air-flush port and operate pump
(air-ilush operation).
Be careful
not
to
damage
air-flush
port
(especially
air-muffler
section) .
Damage to air-ilush port can cause failure.
When
doing
air-flush
operation
Noise level will increase (by
7-8dB).
Install
pump
in an area
which
is
not
exposed
to
debris
such
as
iron
powder,
stone
powder
,
polish
powder
or
wood
dust.
Debris can clog air-muffler, undercutting air-ilush effect.
Be
sure
to
close
isolation
valve between
vacuum
pump
and
vacuum
chamber
during
pump
shut-<lown.
If vacuum pump stops during air-ilush operation, atmospheric air is drawn back from air-ilush port to inside of
vacuum pump, and vacuum on chamber side cannot be maintained. Be sure to close
isolation valve between
vacuum pump and vacuum chamber to prevent the drawback of debris from vacuum pump into vacuum chamber
before stopping vacuum pump.
When
0 eratin with air-ilush OFF closed , 0 erate as
er
4.2.3 a e 23 .
4.2.3 When
returning
to
standard
operation
Before starting air-flush operation, first stop vacuum pump and proceed in
accordance with the following procedure. Never perform this procedure during
operation.
IRemove
air-muffle~
CD
Stop vacuum pump.
® Remove air-muffler from air-flush port.
@ Lightly fit plug to air-flush port with a spanner (nominal dia. 13mm).
*
When
restarting air-flush operation, refer to
4.2.1-4.2.2[page
22 - 23] and
prepare and start.
* Store removed air-muffler and pay attention not to misplace it.
-23-