5. Place the ionizer on the stand and set it up on the side of
the balance.
(Secure the ionizer to the stand using the double-sided
tape if necessary)
1.3. Using the Ionizer
Note: Keep the weighing sample away from the electrodes while neutralizing. Placing the sample too
close to the electrodes may cause the sample to become charged.
Remove any obstacle between the electrodes and the weighing sample.
The electrodes must not be touched with tweezers, etc., while neutralizing.
1. Place the weighing sample to be neutralized in front of the ionizer.
At this time, the weighing sample to be neutralized must be kept in a range of approx. 10 cm to 30 cm
from the electrodes as shown in the figure below.
Neutralization ion generation distribution
Top view
Side view
Electrodes
Stand
Double-sided tape
Weighing sample