9
3. Maintenance of the Electrode Unit
When using the ionizer for a long time, dust and stains may
stick to the electrodes. Clean them periodically to maintain
performance.
When needles of the electrodes are worn down and
neutralization ability is not restored after cleaning them,
replace both electrode units with new ones.
The standard life of an electrode unit is approximately
10,000 hours.
Replacing the Electrode Units
1 Turn the electrode units 45 degrees counterclockwise and remove
them from the ionizer.
2 Replace both electrode units with new ones.
3 Turn the electrode units 45 degrees clockwise and insert them into the
ionizer.
Note: Do not replace, remove and clean the electrode units while the ionizer is operating as there is
danger of electrical shock.
(Replace the balance with the power supply disconnected or with the connection cable
disconnected.)
4. Specifications
Static neutralization method Corona discharge
Effective neutralization range
Approx. 10 cm to 50 cm from the electrode needle
Neutralization performance
1 second (at a distance of approx. 10 cm and 4 kV electrical charge)
Ambient temperature and
humidity
10 to 40 °C, 85% RH or less (no condensation)
Electrode needle Tungsten (with a life of approx. 10,000 hours)
Mass Approx. 300 g (including the stand)
Electrode unit
1
2
Cleaning
Electrode needle