Instruction Manual
1234
To attach
to a wall
Screw
Screw
GC14
Screw
Magnet
Screw
Magnet
Remove 2 screws
Rear
AD8561MI02 / MI04
Multi-Interface
Instruction Manual
1WMPD4004599
Safety Precautions
To prevent accidents due to inappropriate handling, this manual contains the
following warning signs and marks. The meanings of these warning signs and
marks are as follows.
A potentially hazardous situation that could result in death
or serious injury, if not avoided.
A potentially hazardous situation which, if not avoided,
may result in personal injury or property damage.
(1) No parts of this manual may be reproduced without permission. This
manual may not be copied, modified, or translated without the written
permission of A&D Company, Limited.
(2) The contents of this manual are subject to change without notice.
(3) Please contact A&D if you notice any uncertainty, errors, omissions, etc.
in this manual.
(4) A&D Company, Ltd. bears no liability for direct, indirect, special, or
consequential damages due to the operation of this product, even if
advised of the possibility of such damage. Furthermore, A&D assumes no
liability for claims of rights from third parties. Concurrently, A&D assumes
no liability whatsoever for data losses regardless of (3) above.
© 2022 A&D Company, Limited All rights reserved.
ڧ Do not disassemble the interface. It may cause damage or malfunction,
which is not included in the warranty.
ڧ Immediately stop using the interface if an abnormality occurs.
1. Introduction
This manual describes how to handle the Multi-Interface AD8561. Read the
instruction manual thoroughly before use.
For the information on the AD8561 and its related products, please refer to
the A&D website (
https://www.aandd.co.jp/ ).
2. About the AD-8561
2-1. Features
The following operations are enabled when the AD8561 is connected to the
RS-232C output on the GC-series (hereinafter called ‘counting scale’).
ڧ Transmitting and receiving data from Dsub 9 pin
(between the counting scale and a connected device such as a printer)
ڧ Transmitting and receiving data from USB micro B: AD8561MI02 only
(between the counting scale and a connected device such as PC)
ڧ Comparator output: AD8561MI04 only
(from the counting scale to a connected device such as a comparator
lamp)
ڧ Sending texts to the counting scale from USB Type-A with a barcode
reader or keyboard (from a barcode reader/keyboard to the counting
scale)
Each color of the LED lights shown below indicates different status.
LED color
AD-8561 status
Blue Activating
Communication
available
White
A barcode reader or keyboard is connected
(HID setting: key mode)
Communication
available
Yellow
A barcode reader or keyboard is connected
(HID setting: command mode)
Communication
available
Yellow
(blinking)
An input that exceeds 64 digits is received
from a keyboard
(Command mode/Refer to ‘
5-2. USB
Connector
’)
Restart
the device.
Red
(blinking)
Not-supported devices connected to
USB Type-A
Immediately remove
the USB drive
Red Error
Restart
the device.
(For how to restart the device, refer to ‘4-1. Connection to Power Supply’)
2-2. Precautions for Use
ڧ The power for the AD8561 can be supplied from the counting scale with
an accessory cable. When another cable is used, an AC adapter for power
supply may be required. (
Refer to ‘4-1. Connection to Power Supply’)
ڧ Make sure to check the rating of the AC adapter before attaching it.
Inappropriate power supply may cause malfunction or damage to the
internal circuitry.
ڧ Magnets are used for simple fixation.
*When there is any magnetically sensitive device such as a balance nearby, they
can affect the device. In this case, remove the magnets from the AD-8561.
<How to remove the magnets for attaching the interface>
ڧ Magnets are not strong enough to secure the interface to a wall. To attach
to a wall, first attach the AD-8561 to a GC-14, an option for the counting
scale, with screws, and then attach the GC14 to a wall (with screws or by
hanging it on a wall). For the details, refer to the instruction manual for the
GC14.
3. Part Names and Accessories
3-1. Main Unit
3-2. Accessories
ڧ Instruction manual (This manual)
ڧ Cable
(1m, D-sub 9 pin, female-female, for connecting to the counting scale㸧
4. Connection to Other Devices
4-1. Connection to Power Supply
ڧ Supply power from the counting scale by connecting an accessory cable
to the connector on AD8561 (use a Dsub 9 pin).
ڧ When the accessory cable is not used, an AC adapter should be used to
supply power from the DC jack.
z When using a commercially available cable to which the 9
th
pin of
the connector is not connected
z When communicating between the counting scale and AD-8561 at a
distance longer than 1m
*A cable longer than 1m, which is optionally available, is not connected to the
9
th
pin of the connector and therefore requires power supply from an AC
adapter. (Refer to (3) in ‘8-3. Optional Devices’)
ڧ Make sure to check the specifications when using an AC adapter. An AC
adapter with a center negative plug that outputs DC 9 to 12V is usable.
ڧ The main unit is activated when it is connected to the counting scale or
when an AC adapter is inserted into the DC jack. Once it is activated, a
blue LED light is turned on.
ڧ To restart the interface, remove the cable or AC adapter and confirm that
the LED light is turned off. Then, reconnect it to turn the power on.
4-2. Connected Devices
ڧ Set the function F-06-03 of the counting scale to ‘1.’
5. Communication Specifications
5-1. RS-232C Connector (D-sub 9 pin): PRINTER
ڧ Communication between the counting scale and a printer (or PC) is
operable with an RS232C interface.
z Transmission system: EIA RS232C
z Transmission form: Asynchronous, bidirectional, half-duplex
z Data format:
*The settings are changeable (Refer to ’
6. Functions’)
ڧ Pin layout
Baud rate * 2400 bps
Bit length * 7 bit
Parity * EVEN
Start bit 1 bit
Stop bit 1 bit
Code used ASCII
Terminator CRLF (CR: 0Dh, LF: 0Ah)
A
ccessory cable
AD8561
Counting scale
Supply power
LED
Connect to the
counting scale
DC jack
USB connector (Type A)
RS232C connector
Terminal block
(comparator output)
AD8561
MI04
LED
USB connector
(micro B)
DC jack
Connect to the
counting scale
RS232C connector
USB connector (Type A)
AD8561
MI02
AD8561MI02
Dsub (9 pin)
Accessory
cable
USB (micro B)
USB (Type A)
Dsub
(9 pin)
PC
Counting scale
Barcode reade
r
or keyboard
Printer or PC
AD8561MI02 AD8561MI04
Cable with 9
th
pin unconnected
adapter.
(
R
Counting scale
AD8561
Supply power
A
C adapter
AD8561MI04
Dsub (9 pin)
A
ccessory
cable
Terminal block
(4 pin)
Dsub
(9 pin)
or buzzer
Printer or PC
Barcode reade
r
or keyboard
Counting scale
USB (Type A)