Specifications
ATI Model A15/82 Fluoride Monitor with Auto-Calibration
13
O&M Manual
Rev-G (6/10)
REAGENT AND FLUORIDE STANDARD PICKUP TUBING CONNECTION
A special rubber stopper with a ceramic support tube is supplied for the buffer and fluoride standard
bottles. If necessary, adjust the ceramic tube so that it does not hit the bottom of the solution bottles. A length of
pickup tubing is supplied for each of these pickup assemblies with a small filter attached to one end of the buffer
tubing. The buffer tubing is black while the fluoride standard tubing is translucent. After the buffer solution is
ready, feed the end of the tubing opposite the filter up through the ceramic tube. Then insert the tube assembly
into the solution and press the rubber stopper in place. When assembled, the reagent delivery system should
conform to Figure 5 on page 6.
After the reagent bottles have been placed into the brackets or on the mounting panel shelf, attach the
other end of the pickup tubing to the connectors on the reagent pump. The reagent pump has two small barb
connectors on the right side. Press the pickup tube carefully onto each of these barbs, being careful not to pull
on the pump tubing. The tubing from the buffer bottle can be shortened if there is an excess once mounting is
complete.
Figure 7 - Reagent Pickup Tubing Assy