User Manual
Advanced Instruments Inc.
3
2 Quality Control Certification
Date _________ Customer _______________________ Order No.: _____________________ Pass
Model ( ) GPR-2000 Portable O2 Analyzer (A-1163 Battery Assy)
( ) GPR-2000M Portable O2 Analyzer, Flowmeter
( ) GPR-2000CM Portable O2 Analyzer, Coalescing Filter, Flowmeter, Panel
( ) GPR-2000P Portable O2 Analyzer, Integral Pump (A-1157 Battery Assy, A-2166-4)
( ) GPR-2000PC Portable O2 Analyzer, Integral Pump, Coalescing Filter
( ) GPR-2000PI Portable O2 Analyzer, Integral Pump, Intrinsically Safe (A-1158, A-2166-5)
( ) GPR-2000PCI Portable O2 Analyzer, Integral Pump, Coalescing Filter, Intrinsically Safe
( ) GPR-2000PCM Portable O2 Analyzer Integral Pump, Coalescing Filter, Flowmeter, Panel
( ) GPR-2000PCMI Portable O2 Analyzer Integral Pump, Coalescing Filter, Flowmeter,
Panel, Intrinsically Safe (A-1158, A-2166-5)
Sensor
( ) GPR-11-32-RT Oxygen Sensor ( ) XLT-11-24-RT Oxygen Sensor
( ) GPR-11-21-RT Oxygen Sensor
Serial Nos.
Analyzer ______________________________ Sensor __________________________
Accessories Owner’s Manual
( ) PWRS-1002 9VDC Battery Charger/Adapter 110VAC
( ) PWRS-1003 9VDC Battery Charger/Adapter 220VAC
( ) PWRS-1008 9VDC Battery Charger/Adapter 12VDC Auto Cigarette Lighter
CONN-1034 Plug Mini Phone .141 Dia Black Handle
Configuration A-1151-E-2 PCB Assembly
Software version ___________________
Range: 0-1%, 0-5%, 0-10%, 0-25%
Electronics LED indicators: Low battery, charge
Electronic offset
Analog signal output 0-1V
Gas Phase Baseline drift on zero gas < ± 2% FS over 24 hour period on 0-1% range
Noise level < ± 0.5% FS
Span adjustment within 10-50% FS
Final Overall inspection for physical defects
Options
Notes