2 Quality Control Certification
Date: Customer: Order No.: Pass
Model GPR-1000 Portable ppm Oxygen Analyzer
Sensor ( ) GPR-12-100-M ppm Oxygen Sensor
( ) XLT-12-100-M ppm Oxygen Sensor
Serial Nos. Analyzer_________________________ Sensor _____________________________
Accessories Owner’s Manual
( ) PWRS-1002 9VDC Battery Charger/Adapter 110VAC
( ) PWRS-1003 9VDC Battery Charger/Adapter 220VAC
( ) PWRS-1008 9VDC Battery Charger/Adapter 12VDC Auto Cigarette Lighter
CONN-1034 Plug Mini Phone .141 dia. Black Handle
FITN-1003 (3x) Plug Male Quick Disconnect Fittings
TOOL-1001 5/16 Combination Wrench
Configuration A-1151-E-B1 PCB Assembly
Range: 0-1000 ppm, 0-1%, 0-25%
Wetted parts: stainless steel
Electronics Test LED indicators: Low battery, charge
Electronic offset
Analog signal output 0-1V
Gas Phase Test Recovery from air to < 10 ppm in < 1 hour
Baseline drift on zero gas < ± 2% FS over 24 hour period on 0-1% range
Noise level < ± 0.5% FS
Span adjustment within 10-50% FS
Final Overall inspection for physical defects
Options
Notes