Use Instructions
73
8. TECHNICAL SPECIFICATIONS
Equipment type
Internally powered medical electrical (ME) equipment. Type BF Applied
part (IEC60601-1)
The PeriCoach probe is rated to IP27.
Expected Service Life
The expected service life of the PeriCoach probe and charging case is 2
years.
Battery
Type: Lithium-polymer Voltage: 3.7 VDC
Typical service life of the battery: 3 years
Charging Case
Power input: Voltage: 5 VDC, Current: 200 mA
Safety approval requirement for charger or PC providing power to
charging case: IEC60950-1.
Environmental limits
Only store and transport this device in places where the
temperature ranges between -25°C (-13 F) and 70°C
(158 F).
Only use this device when the room temperature is
between 5°C (41 F) and 40°C (104 F).