Rev. 0 | Page 15 of 24
e
nd
ls box
NSOR OFFSET
le
e
ister.
The Alert Manager window also allows the autocycle mode to b
selected for the four uncommitted analog input channels as well
as the two I
SENSE
channels. The desired channels can be selected
by clicking Autocycle Mode in the Alert Manager window a
then selecting the appropriate check boxes in the Channe
on the right side of the window (see Figure 24). Consult the
Register Map window for the assignment of each alert bit into
the three alert registers.
EXTERNAL TEMPERATURE SE
CALIBRATION WINDOW
The External Temperature Sensor Offset Calibration window
is used to calibrate the temperature, correcting errors attributab
to noise that may be coupled onto the Dx+ and Dx− lines of th
remote temperature sensors. Both T
SENSE
1 and T
SENSE
2 can be
offset using the gauges shown in Figure 25. The gauges can be
adjusted from between −32°C and +31.75°C, and the resulting
value is automatically subtracted as a twos complement 8-bit
reading from every temperature measurement before the value
is stored in the relative result reg
07922-024
Fi gure 24. Autocycle Mode in the Alert Manager Window
0
7922-025
Figure 25. External Temperature Sensor Offset Calibration Window