Datasheet
ADM202E/ADM1181A 
Rev. C | Page 10 of 16 
ESD TESTING (IEC1000-4-2) 
IEC1000-4-2 (previously 801-2) specifies compliance testing 
using two coupling methods, contact discharge and air-gap 
discharge. Contact discharge calls for a direct connection to the 
unit being tested. Air-gap discharge uses a higher test voltage, 
but does not make direct contact with the unit being tested. 
With air-gap discharge, the discharge gun is moved toward the 
unit being tested, developing an arc across the air gap. This 
method is influenced by humidity, temperature, barometric 
pressure, distance, and rate of closure of the discharge gun. 
Although less realistic, the contact-discharge method is more 
repeatable and is gaining preference to the air-gap method. 
Although very little energy is contained within an ESD pulse, 
the extremely fast rise time coupled with high voltages can 
cause failures in unprotected semiconductors. Catastrophic 
destruction can occur immediately as a result of arcing or 
heating. Even if catastrophic failure does not occur immediately, 
the device might suffer from parametric degradation, which can 
result in degraded performance. The cumulative effects of 
continuous exposure can eventually lead to complete failure. 
I/O lines are particularly vulnerable to ESD damage. Simply 
touching or plugging in an I/O cable can result in a static 
discharge, which can damage or completely destroy the 
interface product connected to the I/O port. Traditional ESD 
test methods, such as the MIL-STD-883B method 3015.7, do 
not fully test a product’s susceptibility to this type of discharge. 
This test was intended to test a product’s susceptibility to ESD 
damage during handling. Each pin is tested with respect to all 
other pins. There are some important differences between the 
traditional test and the IEC test: 
•  The IEC test is much more stringent in terms of discharge 
energy. The injected peak current is over four times greater. 
•  The current rise time is significantly faster in the IEC test. 
•  The IEC test is carried out while power is applied to 
the device. 
It is possible that the ESD discharge could induce latch-up in the 
device being tested. Therefore, this test is more representative of a 
real-world I/O discharge where the equipment is operating 
normally with power applied. For peace of mind, however, both 
tests should be performed to ensure maximum protection 
during both handling and field service. 
R1 R2
C1
DEVICE
UNDER TEST
HIGH
VOLTAGE
GENERATOR
ESD TEST METHOD R2 C1
H. BODY MIL-STD883B 1.5kΩ 100pF
IEC1000-4-2 330Ω 150pF
00066-018
Figure 18. ESD Test Standards 
100
I
PEAK
 (%)
90
36.8
10
 t
DL
t
RL
TIME t
00066-019
Figure 19. Human Body Model ESD Current Waveform 
100
I
PEAK
 (%)
90
10
TIME t
 30ns
 60ns
0.1 TO 1ns
00066-020
Figure 20. IEC1000-4-2 ESD Current Waveform 
The ADM202E/ADM1181E products are tested using both of 
the previously mentioned test methods. Pins are tested with 
respect to all other pins as per the MIL-STD-883B specification. 
In addition, I/O pins are tested as per the IEC test specification. 
The products were tested under the following conditions: 
•  Power-On 
•  Power-Off 
There are four levels of compliance defined by IEC1000-4-2. The 
ADM202E/ADM1181A products meet the most stringent level 
of compliance both for contact and for air-gap discharge. This 
means that the products are able to withstand contact discharges 
in excess of 8 kV and air-gap discharges in excess of 15 kV. 










