Use Instructions

Table Of Contents
35
Version C
Clinician Instructions for Use Device and Mobile Application
Figure 23. Adhesive storage compartment in the Base Station.
10.
PATIENT MODE MEASUREMENTS
10.1 Measurement preparation
1)
Ensure that the room is quiet (e.g. no talking, no loud machines in the background such as air
conditioners, fans, radios, TVs, etc.).
2)
The patient should
remove their shirt and/or bra
to expose the chest. For men: trim chest hair if needed.
3)
Remove the CPM Device from the Base Station, checking that the device charge LEDs on the device and
Base Station are green or yellow. Replace the adhesives if needed (instructions on page 37).
4)
Peel off the two clear liners to expose the skin adhesive and then place them back into the Base Station,
aligning their shapes to the appropriate wells.
5)
Place the device as designated by the clinician (guidelines given on page 17), pressing down around all the
sensors. Use the alignment tool (use is described on page 33) and/or Base Station mirror to help, if needed.
10.2 Patient mode workflow
IMPORTANT
For best results, take measurements at the same time each day (recommended at morning wake-up).
IMPORTANT
For best results, apply the device as consistently as possible during each use, in the location used to take
the baseline reading.
IMPORTANT
Stay stationary and do not speak for the duration of the measurement.