Datasheet
Data Sheet ADAS1000/ADAS1000-1/ADAS1000-2
Rev. A | Page 67 of 80
Table 45. Lead-Off Status Register (LOFF) Address 0x1D, Reset Value = 0x000000
R
/W
Default Bit Name Function
R 0 23 RLD lead-off
Status
Electrode connection status.
If either dc or ac lead-off is enabled, these bits are the corresponding lead-off status. If
both dc and ac lead-off are enabled, these bits reflect only the ac lead-off status. DC
lead-off is available in the DCLEAD-OFF register (see Table 46).
The common electrodes have only dc lead-off detection.
An ac lead-off signal can be injected into the common electrode, but there is no ADC
input to measure its amplitude. If the common electrode is off, it affects the ac lead-off
amplitude of the other electrodes.
These bits accumulate in the frame buffer and are cleared when the frame buffer is
loaded into the SPI buffer.
0 = electrode is connected.
1 = electrode is disconnected.
22 LA lead-off status
21
LL lead-off status
20 RA lead-off status
19 V1 lead-off status
18
V2 lead-off status
13 CELO
R
0
[17:14]
Reserved
Reserved.
R 0 12 LAADCOR ADC out of range error.
These status bits indicate the resulting ADC code is out of range.
These bits accumulate in the frame buffer and are cleared when the frame buffer is
loaded into the SPI buffer.
11 LLADCOR
10 RAADCOR
9 V1ADCOR
8 V2ADCOR
R 0 [7:0] Reserved Reserved.
Table 46. DC Lead-Off Register (DCLEAD-OFF) Address 0x1E, Reset Value = 0x000000
1
R
/W
Default Bit Name Function
R 0 23 RLD input
overrange
The dc lead-off detection is comparator based and compares to a fixed level. Individual
electrode bits flag indicate if the dc lead-off comparator threshold level has been
exceeded.
0 = electrode < overrange threshold, 2.4 V.
1 = electrode > overrange threshold, 2.4 V.
22 LA input
overrange
21 LL input
overrange
20 RA input
overrange
13 CE input
overrange
R 0 [19:14]
[8:3]
Reserved Reserved.
R 0 12 RLD input
underrange
The dc lead-off detection is comparator based and compares to a fixed level. Individual
electrode bits indicate if the dc lead-off comparator threshold level has been exceeded.
0 = electrode > underrange threshold, 0.2 V.
1 = electrode < underrange threshold, 0.2 V.
11 LA input
underrange
10 LL input
underrange
9 RA input
underrange
2
CE input
underrange
R 0 [1:0] Reserved
1
This register is not part of framing data, but can be read by issuing a register read command of this address.