Datasheet
Data Sheet ADAS1000/ADAS1000-1/ADAS1000-2
Rev. A | Page 63 of 80
Table 35. Frame Control Register (FRMCTL) Address 0x0A, Reset Value = 0x079000
R/W
Default Bit Name Function
R/W 0 23 LEAD I/LADIS Include/exclude word from ECG data frame. If the electrode/lead is included in the data-
word and the electrode falls off, the data-word is undefined.
0 (default)
= included in frame.
1 = exclude from frame.
R/W 0 22 LEADII/LLDIS
R/W 0 21 LEADIII/RADIS
R/W 0 20 V1DIS
R/W 0 19 V2DIS
R/W
1111
[18:15]
Reserved
Reserved, set to 1111.
R/W 0 14 PACEDIS
1
Pace detection.
0 (default)
= included in frame.
1 = exclude from frame.
R/W 0 13 RESPMDIS
1
Respiration magnitude.
0 (default)
= included in frame.
1 = exclude from frame.
R/W 1 12 RESPPHDIS
1
Respiration phase.
0 = included in frame.
1 (default)
= exclude from frame.
R/W 0 11 LOFFDIS Lead-off status.
0 (default)
= included in frame.
1 = exclude from frame.
R/W 0 10 GPIODIS GPIO word disable.
0 (default)
= included in frame.
1 = exclude from frame.
R/W 0 9 CRCDIS CRC word disable.
0 (default)
= included in frame.
1 = exclude from frame.
R/W 0 8 RESERVED Reserved, set to 0.
R/W 0 7 ADIS Automatically excludes PACEDIS[14], RESPMDIS[13], LOFFDIS[11] words if their flags are
not set in the header.
0 (default)
= fixed frame format.
1 = autodisable words (words per frame changes).
R/W 0 6 RDYRPT Ready repeat. If this bit is set and the frame header indicates data is not ready, the frame
header is continuously sent until data is ready.
0 (default)
= always send entire frame.
1 = repeat frame header until ready.
R/W
0
5
Reserved
Reserved, set to 0 .
R/W 0 4 DATAFMT Sets the output data format, see Figure 57.
0 (default) = digital lead/vector format (available only in 2 kHz and 16 kHz data rates).
1 = electrode format.
R/W 00 [3:2] SKIP[1:0] Skip interval. This field provides a way to decimate the data.
00 (default) = output every frame.
01 = output every other frame
1× = output every 4
th
frame.
R/W 00 [1:0] FRMRATE[1:0] Sets the output data rate.
00 (default) = 2 kHz output data rate.
01 = 16 kHz output data rate.
10 = 128 kHz output data rate (DATAFMT must be set to 1).
11 = 31.25 Hz.
1