Datasheet

AD7984
Rev. A | Page 4 of 24
VDD = 2.5 V, VIO = 2.3 V to 5.5 V, REF = 5 V, T
A
= −40°C to +85°C, unless otherwise noted.
Table 3.
Parameter Conditions Min Typ Max Unit
REFERENCE
Voltage Range 2.9 5.1 V
Load Current 1.33 MSPS 520 μA
SAMPLING DYNAMICS
−3 dB Input Bandwidth 10 MHz
Aperture Delay 2 ns
DIGITAL INPUTS
Logic Levels
V
IL
VIO > 3 V –0.3 +0.3 × VIO V
V
IH
VIO > 3 V 0.7 × VIO VIO + 0.3 V
V
IL
VIO ≤ 3 V –0.3 +0.1 × VIO V
V
IH
VIO ≤ 3 V 0.9 × VIO VIO + 0.3 V
I
IL
−1 +1 μA
I
IH
−1 +1 μA
DIGITAL OUTPUTS
Data Format Serial 18 bits, twos complement
Pipeline Delay
Conversion results available immediately
after completed conversion
V
OL
I
SINK
= +500 μA 0.4 V
V
OH
I
SOURCE
= −500 μA VIO − 0.3 V
POWER SUPPLIES
VDD 2.375 2.5 2.625 V
VIO Specified performance 2.3 5.5 V
VIO Range 1.8 5.5 V
Standby Current
1, 2
VDD and VIO = 2.5 V 1.1 mA
Power Dissipation 1.33 MSPS throughput 10.5 14 mW
Energy per Conversion 7.9 nJ/sample
TEMPERATURE RANGE
3
Specified Performance T
MIN
to T
MAX
−40 +85 °C