Datasheet

AD7194 Data Sheet
Rev. A | Page 42 of 56
The output data rate equals
f
ADC
= 1/t
SETTLE
= f
CLK
/(3 × 1024 × FS[9:0])
where:
f
ADC
is the output data rate.
f
CLK
is the master clock (4.92 MHz nominal).
FS[9:0] is the decimal equivalent of Bit FS9 to Bit FS0 in the
mode register.
When the analog input is constant or a channel change occurs,
valid conversions are available at a constant output data rate.
When conversions are being performed on a single channel and
a step change occurs on the analog input, the ADC continues to
output fully settled conversions if the step change is synchronized
with the conversion process. If the step change is asynchronous,
one conversion is output from the ADC that is not completely
settled (see Figure 38).
ANALOG
INPUT
ADC
OUTPUT
FULLY
SETTLED
1/
f
ADC
08566-037
Figure 38. Sinc
3
Zero Latency Operation
Table 31 provides examples of output data rates and the corres-
ponding FS values.
Table 31. Examples of Output Data Rates and the
Corresponding Settling Time (Zero Latency)
FS[9:0] Output Data Rate (Hz) Settling Time (ms)
480 3.3 300
96 16.7 60
80 20 50
Sinc
3
50 Hz/60 Hz Rejection
Figure 39 show the frequency response of the sinc
3
filter when
FS[9:0] is set to 96 and the master clock equals 4.92 MHz. The
output data rate is equal to 50 Hz when zero latency is disabled
and 16.7 Hz when zero latency is enabled. The sinc
3
filter gives
50 Hz ± 1 Hz rejection of 95 dB minimum for a stable master clock.
–120
–110
–100
–90
–80
–70
–60
–50
–40
–30
–20
–10
0
0 25 50
75
100
125
150
FREQUENCY (Hz)
FILTER GAIN (dB)
08566-038
Figure 39. Sinc
3
Filter Response (FS[9:0] = 96)
When FS[9:0] is set to 80 and the master clock equals
4.92 MHz, 60 Hz rejection is achieved (see Figure 40). The
output data rate is equal to 60 Hz when zero latency is disabled
and 20 Hz when zero latency is enabled. The sinc
3
filter has
rejection of 95 dB minimum at 60 Hz ± 1 Hz, assuming a stable
master clock.
–120
–110
–100
–90
–80
–70
–60
–50
–40
–30
–20
–10
0
0 30 60 90 120 150
FREQUENCY (Hz)
FIL
TER GAIN (dB)
08566-039
Figure 40. Sinc
3
Filter Response (FS[9:0] = 80)