Datasheet
Data Sheet AD633
Rev. J | Page 3 of 20
SPECIFICATIONS
T
A
= 25°C, V
S
= ±15 V, R
L
≥ 2 kΩ.
Table 1.
AD633J, AD633A
Parameter Conditions Min Typ Max Unit
TRANSFER FUNCTION
( )( )
Z
Y2Y1X2X1
W +
−−
=
V10
MULTIPLIER PERFORMANCE
Total Error −10 V ≤ X, Y ≤ +10 V ±1 ±2
1
% full scale
T
MIN
to T
MAX
±3
% full scale
Scale Voltage Error SF = 10.00 V nominal ±0.25% % full scale
Supply Rejection V
S
= ±14 V to ±16 V ±0.01 % full scale
Nonlinearity, X X = ±10 V, Y = +10 V ±0.4 ±1
1
% full scale
Nonlinearity, Y Y = ±10 V, X = +10 V ±0.1 ±0.4
1
% full scale
X Feedthrough Y nulled, X = ±10 V ±0.3 ±1
1
% full scale
Y Feedthrough X nulled, Y = ±10 V ±0.1 ±0.4
1
% full scale
Output Offset Voltage
2
±5 ±50
1
mV
DYNAMICS
Small Signal Bandwidth V
O
= 0.1 V rms 1 MHz
Slew Rate V
O
= 20 V p-p 20 V/µs
Settling Time to 1% ΔV
O
= 20 V 2 µs
OUTPUT NOISE
Spectral Density 0.8 µV/√Hz
Wideband Noise f = 10 Hz to 5 MHz 1 mV rms
f = 10 Hz to 10 kHz 90 µV rms
OUTPUT
Output Voltage Swing ±11
1
V
Short Circuit Current R
L
= 0 Ω 30 40
1
mA
INPUT AMPLIFIERS
Signal Voltage Range Differential ±10
1
V
Common mode ±10
1
V
Offset Voltage (X, Y) ±5 ±30
1
mV
CMRR (X, Y) V
CM
= ±10 V, f = 50 Hz 60
1
80 dB
Bias Current (X, Y, Z) 0.8 2.0
1
µA
Differential Resistance 10 MΩ
POWER SUPPLY
Supply Voltage
Rated Performance ±15 V
Operating Range ±8
1
±18
1
V
Supply Current Quiescent 4 6
1
mA
1
This specification was tested on all production units at electrical test. Results from those tests are used to calculate outgoing quality levels. All minimum and maximum
specifications are guaranteed; however, only this specification was tested on all production units.
2
Allow approximately 0.5 ms for settling following power on.