Product data

30
www.rosemount.com
Rosemount 5300 Series
October 2014
Temperature and pressure limits
Ambient Temperature
The maximum and minimum ambient temperature for the electronics depends on the process
temperature (as described by the graph below) and on the approval (see “Product
Certifications” on page 55).
Note
Nozzle insulation for the HTHP version (Operating Temperature and Pressure code H) should
not exceed 4 in. (10 cm).
Note
The temperature range for the optional Integral Display is -4 °F to 158 °F (-20 °C to 70 °C).
Note
In applications where the ambient temperature exceeds the limits of the electronics, a Remote
Mounting connection can be used. The maximum temperature for the Remote Mounting
connection at the vessel connection point is 302 °F (150 °C).
Storage Temperature
-58 °F to 194 °F (-50 °C to 90 °C)
With Integral Display: -40 °F to 185 °F (-40 °C to 85 °C)
Ambient Temperature °F (°C)
Process Temperature °F (°C)
185 (85)
131 (55)
100 (38)
50 (10)
-40 (-40)
392 (200) 752 (400)
-17 (-27)
-40 (-40)
200 (93) 400 (204) 600 (316) 800 (427)
-320 (-196)