WARNING:
Risk of Suffocation — Dispose of the packaging material, observing the
applicable waste-control regulations and keeping it out of children's reach.
CAUTION:
Personal Injury — All devices used must comply with the applicable IEC/EN or
ISO standard (PCs, for instance, must comply with IEC 60950). This applies to
custom designs in particular.
During operation from flat-screen display (contact to monitor) do not
simultaneous touch patient.
CAUTION:
Personal Injury, Equipment Damage — Check that the ambient conditions
specified by the manufacturers of the different devices are ensured.
Before cleaning the flat-screen display, turn it off and remove power cord from
power supply. Do not spray liquid cleaning agents directly onto the display.
Spray the cleaning agent onto a cloth and wipe the display clean without
exerting any pressure.
Do not place any objects on the signal cables. Route the signal cables and the
power cable to the flat-screen display such that they do not present a
stumbling hazard.
CAUTION:
Equipment Damage — Do not set up the flat-screen display in the immediate
vicinity of a window. Rain, water, humidity and sunlight may damage the flat-
screen display.
Do not exert pressure on the LCD. Excessive pressure may permanently
damage the display.
For safe storage of the flat-screen display, a temperature between -20°C and
+60°C must be ensured. Temperatures above or below this range may
permanently damage the flat-screen display. Always store your flat-screen
display in its original shipping box.
WARNING:
Personal Injury — During intracardiac application, it is recommended to
connect the individual devices to the room’s potential equalization system via
separate potential equalization cables.
For intracardiac application, the application guidelines for medical electrical
devices set forth in the German standard VDE 0753, part 2, must be observed.