Specifications

Users Manual: MD-CP42AUPG-MED
© 2011 Richardson Electronics GmbH
Division Canvys
24
Instruction set for serial communication with RS-232
Baud rate: 9600 (bps)
Data bit: 8 bit
Parity: none
Stop bit: 1 bit
Note: All commandos are made by combination of numeric value which is expressed by
hexadecimal format. Like as “0A”, “FF”.
Note: All communication needs a handshake replay.
1.0 Control Read command:
command format: 04 03 ID CK1
reply command format: 04 ID DD CK2
CK1:checksum for checking command weather valid. CK1=100-04-03-ID
CK2:checksum for checking command weather valid. CK2=100-04-ID-DD
1.1 Color Mode status (ID=0x01)
command 04 03 01 CK1
reply: 04 01 00 CK2 - Gamma OFF mode.
reply: 04 01 01 CK2 - Gamma2.0 Mode.
reply: 04 01 02 CK2 - Gamma2.2 Mode.
reply: 04 01 03 CK2 - Gamma1.8 Mode.
reply: 04 01 04 CK2 - DICOM Mode.
reply: 04 01 05 CK2 - Calibration Mode.
1.2 Color temperature status. (ID=0x03)
command 04 03 03 CK1
reply: 04 03 00 CK2 - Cool
reply: 04 03 01 CK2 - Neutral
reply: 04 03 02 CK2 - Warm
reply: 04 03 03 CK2 - User
1.3 Source input status.(ID=0x04)
command 04 03 04 CK1
reply: 04 04 00 CK2 - DVI-I Analog Input.
reply: 04 04 01 CK2 - DVI-I Digital Input.
reply: 04 04 02 CK2 - DVI-D Digital Input.
reply: 04 04 03 CK2 - S-Video Input.
reply: 04 04 04 CK2 - Composite Input.
reply: 04 04 05 CK2 - YPbPr Input.
reply: 04 04 06 CK2 - RGBS Input.
reply: 04 04 07 CK2 - SDI Input.
1.4 Brightness status (ID=0x10)
command 04 03 10 CK1
reply: 04 10 DD CK2 - Brightness value(from 0x00-0x64)
1.5 Contrast status. (ID=0x11)
command 04 03 11 CK1
reply: 04 11 DD CK2 - Contrast value(from 0x00-0x64)