MAINTENANCE 43
8.8 Replacing the fuse
TO BE CARRIED OUT BY A QUALIFIED ELECTRICIAN ONLY.
The main power supply incorporates a fuse protection on both the live and the
neutral connections, the locations of the fuses are shown in Figure 2 on page 8.
The main connection plug on the 230v instrument should also be fitted with a
fuse.
The appropriate fuse ratings are as follows:-
Voltage Harrier 15/80
(air cooled)
Harrier 18/80
Refrigerated
120v 15 amp (6.3 x 32mm) 15 amp (6.3 x 32mm)
230v 5 amp (5 x 20mm) 10 amp (6.3 x 32mm)
230v plug 7 amp (1/4” x 1”) 13 amp (1/4” x 1”)
If one or both of the fuses have blown, and the centrifuge is switched on, the
display will not function.
To replace the centrifuge fuse proceed as follows:-
1. Disconnect the power supply at the main plug.
2. Move the centrifuge to gain access to the rear,
3. Release the fuses using a flat head screwdriver. Inspect and replace as
necessary with the correct rating. For 230v model, the plug fuse may also
need to be checked.
NB - If the fuse blows immediately or blows again during normal use, the power
supply should be disconnected and the service engineer called.
9. Product Disposal - ISO 14001 Compliance
This product should be treated as industrial waste and disposed of accordingly.
There are no toxic material used in the manufacture of this product. The
majority of materials used in this product are recyclable, and all can be
disposed of safely. Where the product has refrigeration, it is important that prior
to disposal the refrigerant gas is recovered by a qualified person. The insulation
material is non-toxic but could be an irritant. If removed from the product it
should be bagged and disposed of at an authorised site.