Instruction manual

ERROR MESSAGE TABLES 28
ERROR MESSAGE EXPLANATION EFFECT ON MACHINE OPERATOR ACTION
OVER 45° Bowl temperature is over 45°C
Machine will stop 1. Press CE and open lid
2. Remove rotor
3. Disconnect the power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
BAD THERM Malfunction of temperature
sensor
Machine will stop
Error cannot be cancelled until rotor has
stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect the power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
HOT COMP Compressor has over-heated Machine will stop
Error cannot be cancelled until rotor has
stopped turning
1. Press CE and open lid
2. Ensure that the air filter and inlet
vents are not blocked
TEMP LOW Set value has been exceeded Machine will stop
Error cannot be cancelled until rotor has
stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect the power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
TEMP HIGH Set value has not been
achieved
Machine will stop
Error cannot be cancelled until rotor has
stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect the power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER