Instruction manual

ERROR MESSAGE TABLES 27
ERROR MESSAGE EXPLANATION EFFECT ON MACHINE OPERATOR ACTION
BAD CHKS Link to inverter is not
functioning
Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect from power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
BAD MEM Software memory problem Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect from power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
FRAME Software data not correctly
synchronised
Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect from power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
LID OPEN Lid is open during Pre-cool Machine will keep running 1. Close lid
2. Press CE