Instruction manual

ERROR MESSAGE TABLES 26
ERROR MESSAGE EXPLANATION EFFECT ON MACHINE OPERATOR ACTION
NO TACHO Fault in tacho signal Machine will stop
Lid lock will prevent normal lid operation
1. Wait for 5 minutes or until the rotor
has stopped turning
2. Follow the procedure on page 22 for
opening the lid
3. Remove the rotor
4. Disconnect the power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
BAD STOP Rotor is stationary when it
should be running
Machine is stationary Press CE
POW FAIL There has been an
interruption of the power
supply to the machine
Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
1. Press CE and open lid
2. Check power supply
SER TOUT Inverter is not responding to
the software
Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect from power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER