Instruction manual
ERROR MESSAGE TABLES 25
ERROR MESSAGE EXPLANATION EFFECT ON MACHINE OPERATOR ACTION
UNLOCKED Malfunction of lid lock
mechanism or interlock switch
Machine will stop
Lid lock will prevent normal lid operation
1. Open lid (if lid is locked follow
procedure on page 22)
2. Remove rotor
3. Disconnect from power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
SPEED +-.
Set value has been exceeded Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect the power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
SPEED - Set value has not been
achieved
Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
1. Press CE and open lid
2. Remove rotor
3. Disconnect the power supply
DO NOT USE THE MACHINE UNTIL IT
HAS BEEN CHECKED BY A SERVICE
ENGINEER
HOT MOTOR Motor has overheated Machine will stop
Error cannot be cancelled until the rotor
has stopped turning
Thermostat will reset once the motor has
cooled
1. Press CE and open lid
2. Ensure that the inlet and outlet vents
are not blocked