Instruction manual

GOOD OPERATOR PRACTICE 13
2.1 Connecting the power supply
230v centrifuge - Connect the 3-core cable to a 3-pin plug, fitted with a 13 amp
fuse.
Brown wire to Live (L) terminal
Blue wire to Neutral (N) terminal
Yellow/Green wire to Earth (E) terminal
NOTE: BS EN 61010-2-020 ‘Particular Requirements for Centrifuges’ states
that a remote switch, preferably adjacent to the room exit is a requirement in
case of unit malfunction.
120v centrifuge -
supplied with fitted plug.
2.2 Centrifuge restraint
A suggested restraint system (part number SG.66500.1214), which secures the
centrifuge to a sound work surface, is shown in Figure 6. It consists of 4 anchor
brackets, positioned at each corner of the centrifuge base, which are bolted or
screwed to the work surface.
Figure 6 Suggested Restraint System
2.3 Accessories supplied with each centrifuge
Description Quantity Part No.
Emergency over-ride key 1 76600.004
Guarantee card 120v/230v 1 71100.1052/958
Operating manual 1 71100.1378
Box spanner 1 96500.384AD