Instruction manual

INTRODUCTION TO YOUR CENTRIFUGE 10
1.1 Specifications
Refrigerated Non Refrigerated
Power supply options
120v 60Hz single phase
230v 50Hz single phase
120v 50/60Hz single phase
230v 50/60Hz single phase
Power consumption
at full load
(Acceleration power
shown in brackets)
120v - 2000 watts (3000 watts)
230v - 800 watts (1600 watts)
120v - 1020 watts (1500 watts)
230v - 420 watts (1050 watts)
Time DISPLAY/RANGE
0 - 59 minutes 59 seconds
1 hour - 99 hours 59 minutes
or Time Hold
1 hour - 99 hours 59 minutes
or Time Hold
Timing Accuracy
0.1 second
0.1 second
Speed range
up to 18,000 rpm
up to 15,000 rpm
Speed control accuracy
± 10 rpm
± 10 rpm
Speed DISPLAY
± 20 rpm
± 20 rpm
Speed SETTING increments
1 rpm
1 rpm
RCF range
up to 28975g up to 20154g
RCF SETTING increments
1g 1g
Temp range
0 to 40°C*
_
Temp accuracy
± 2°C
_
Pre cool
-9 to ambient
_
Acceleration rates
10 off
10 off
Brake rates
10 off
10 off
Dimensions
Height 340 mm
Width 670 mm
Depth 580 mm
(not including mains cable)
Weight 65 kg
Height 340 mm
Width 470 mm
Depth 580 mm
(not including mains cable)
Weight 41 kg
Standard operating
conditions
Atmospheric pressure
950-1015 mbar
(13.78 - 14.71 psi)
Ambient temperature
+1C to +35°C
Max. operating temperature
+3C
Humidity
up to 70% RH
Atmospheric pressure
950-1015 mbar
(13.78 - 14.71 psi)
Ambient temperature
+1C to +35°C
Max. operating temperature
+3C
Humidity
up to 70% RH
*Temperatures lower than 0°C can be obtained, dependent on rotor, at maximum speed.