User Manual
17
150
Series
T
roubleshooting
Common Issues (Continued) 
Symptom  Cause  Remedy 
OPTICAL ISSUES
The color of the image is
not accurate 
The brightness adjustment knob is not in the right position  Adjust the brightness knob to a higher or lower setting for color
clarity 
No filter is used or filter is in use  Remove color filter if natural light is desired, or insert desired
filter 
MECHANICAL ISSUES
The objective touches the
cover slip 
The cover slip is not standard sized  Replace the cover slip with the appropriate 0.17mm thickness slip 
The limit-stop is set too high or not engaged  Be careful to avoid contact between objective and the slide when
the limit stop is not engaged. To reengage, focus the sample, then
lock the limit stop into place to set max height at a safe but us-
able distance. 
Unable to move the slide
smoothly 
The slide is not secured correctly  Adjust the slide to use the stage clips and secure the sample 
The mechanical stage is not properly secured  Tighten the mechanical stage screws to better secure the stage 
Focus knob does not turn 
The tension knob is too tight  Loosen it  by adjusting  the  tension  ring inside the  coarse  focus
knob counterclockwise (close  to  the arm of the  microscope  on
the left of the microscope) 
Stage declines by itself 
The tension knob is too lose  Tighten it by adjusting  the tension ring inside the coarse  focus
knob clockwise  (close  to the arm of the microscope  on the left
of the microscope) 
The focusing knob won’t
raise the stage 
Limit-stop is engaged  Disengage the limit stop on the rear of the stage of the micro-
scope 
ELECTRICAL ISSUES
The bulb/light source
flickers 
The bulb is close to burning out  Please contact us in regards to this issue. LED lights do not burn
out, so there may be another electrical issue 
The microscope does not
light up 
The microscope is unplugged  Insert the plug into the wall socket to achieve electrical illumina-
tion 