Specification Sheet

www.Amprobe.com
Data Sheet
Electrical Specification Accuracy (23 °C ± 5 °C & < 75% R.H.), cont.
ACD-14 PLUS 600A Clamp-On Multimeter with Dual Display
Function Range Accuracy
Max/Hold (AC Current only) Specified accuracy ± 50 digits for changes > 25ms in duration
Ohms
400.0 ± (0.8% rdg + 8 digits)
4.000k, 40.00k, 400.0k ± (0.6% rdg + 4 digits)
4.000M ± (1.0% rdg + 4 digits)
40.00M ± (2.0% rdg + 4 digits)
Open Circuit Voltage: 0.4VDC typical
Audible Continuity Tester Audible threshold: between 10 and 120.
Capacitance
500.0nF, 5.000µF, 50.00µF,
500.F, 3000µF* ± (3.5% rdg + 6 digits)***
*Additional 50.00nF range accuracy is not specified
**Accuracies with film capacitor or better
***Specified with battery voltage above 2.8V (approximately half full battery). Accuracy decreases gradually to 12% at low battery
warning voltage of approximately 2.4V
DC µA
Range Accuracy Burden Voltage
400.0µA ± (2.0% rdg + 4 digits) 2.8mV/µA
200A ± (1.2% rdg + 3 digits) 2.8mV/µA
AC µA (50Hz - 500Hz)
Range Accuracy Burden Voltage
400.0µA ± (2.0% rdg + 5 digits) 2.8mV/µA
200A ± (1.5% rdg + 5 digits ) 2.8mV/µA
Frequency ( Voltage input only)
Volts Range Sensitivity Accuracy
400.0mV 10Hz to 2kHz 300mV ± (0.5% rdg + 4 digits
4.000V 5Hz to 40kHz 4V ± (0.5% rdg + 4 digits)
40.00V 5Hz to 100kHz 28V ± (0.5% rdg + 4 digits)
400.0V 5Hz to 100kHz 100V ± (0.5% rdg + 4 digits)
600V 5Hz to 5kHz 500V ± (0.5% rdg + 4 digits)
Diode Tester
Open Circuit Voltage Test Current (Typical)
< 1.6 VDC 0.25mA
Temperature (Type-K Thermocouple)
Range Accuracy*
-20°C to 300°C ± (2% rdg + 3 °C)
301 °C ~ 537 °C ± (3% rdg + 3 °C)
-F to 572°F ± (2% rdg + 6 °F)
573 °F ~ 999 °F ± (3% rdg + 6 °F)
*Type-K thermocouple range & accuracy not included