Service manual

MDS E-SCAN MONITOR OPERATION
43
Note: The readings below are just an example, they will vary from application to application.
5.2.4 The Menu Options in Normal Mode
Normal mode is entered by default, upon power up. To gain access to other system features
select the MENU button on the Front Panel Touch Pad.
When the MENU button is
selected, these menu options are
displayed on the normal mode
screen:
RESET:
Option resets the Monitor console
and clears all error conditions. It
performs the same function as the
reset button located on the door of
the MDS E-Scan cabinet. All
errors require manual reset as per
BLRBAC standards.
LOGS:
Option allows the operator to view historical data pertaining to system errors and warnings,
purge cycle, divert conditions, and comparator readings.
CONFIG:
Option allows the operator to change system parameters. This consists of state conditions such
as trending, as well as configuration for a one or two unit divert. In addition, the operator can
enter set points for all system limits, such as trending, comparator, and temperature limits.
DIAG: (DIAGNOSTICS)
Option provides the operator with data that can be used to test and troubleshoot system problems
(see Diagnostics Mode Selection). The operator can display all monitored voltages, test relays
on the interface board, and enter system time and date in cases of battery failure.
PRINT:
Option allows the operator to dump all system logs to the printer.
Normal Mode Screen With Menu Options