MDS E-SCAN MONITOR OPERATION
42
SYSTEM:
This status field is used to monitor many system functions and to issue warnings instead of errors
so that the operator can be notified of problems that are not fatal but need to be addressed. The
following warnings can be displayed in this status field:
CHK LAMP VOLTAGE: This indicates the lamp voltage in a sensing head is above 4.0
volts. CAUTION- If lamp voltage reaches 4.2 volts the
refractometer will be taken out of service. To adjust see
Maintenance section 8.5.
CHK PURGE LOG: This indicates that a second or third attempt was needed to clean
the prism. Therefore, the operator should check the purge log in
order to determine how much steam time was needed during the
purge operation.
CHK ANLG. SOLIDS: This indicates that the solids reading derived by the 0-10 volt input
to the Monitor unit does not agree with the solids reading sent to
the Monitor unit via the serial port.
CHK ANLG. TEMP.: This indicates that the temperature reading derived by the
temperature voltage input to the Monitor unit does not agree
with the temperature reading sent to the Monitor unit via the
serial port.
CHK U1-U2 COM2: This indicates that the communication between the MPR E-Scan
units via serial port 2 has failed, and therefore all communication
between the MPR E-Scan units is now going through the Monitor
unit.
PRINTER ERROR: This indicates that there is a problem with the printer.
TREND:
This status field is used to monitor the rate of change in solids readings in order to predict how
much time would have to pass at the present rate of change until a divert occurred. This is also a
warning status, and is triggered by the trend limit specified by the operator in the Monitor
Configuration Limits menu option. For example, if the trend limit is set for 30 minutes, then a
trend warning will only be activated if a divert is anticipated within 30 minutes.
DIVERT ERROR: This indicates a divert has occurred. The cause could be any of the
following which would be specified by the printout: Low solids
Divert, Unit Failure Divert, Remote Divert, Diagnostic Divert.