Service manual

MDS E-SCAN MONITOR OPERATION
41
PURGE MISSED: This indicates that the steam solenoid relay was not activated
during the purge operation and therefore the purge did not occur.
PURGE INVALID: This indicates that either the steam solenoid relay was activated
during a non-purging operation or there is an open-line condition.
PURGE NOT ACTIVE: This is caused by not having purge cycle times set. This alarm
flashes above the alarm box and does not take the refractometer off
line.
ISO.VALVE:
This status field monitors the Isolation Valve in order to indicate valve open or valve closed
status.
VALVE CLOSED: This indicates that the isolation valve is closed.
5.2.3 Monitor Status Box MDS in Normal Mode
P/SUPPLY:
This status field monitors the Monitor +15 and -15 volt power supply lines. The following errors
can be displayed in this status field:
+15 VOLT ERROR: This indicates that the plus 15 volt supply is outside the allowable
10% tolerance levels which means it is either below +13.5 volts or
above +16.5 volts.
-15 VOLT ERROR: This indicates that the minus 15 volt supply is outside the
allowable 10% tolerance levels which means it is either above
-13.5 volts or below -16.5 volts.
COMPARATOR:
This status field monitors the readings from both MPR E-Scan units and is used to indicate an
error if the difference in the solids readings exceeds the comparator limit. The operator has the
capability of changing the comparator limit to a value between 1% and 2% absolute solids by
using the Monitor Configuration Limits menu option.
COMPARATOR ERROR: The difference in solids readings exceeds the comparator limit.