MDS E-SCAN MONITOR OPERATION
40
+15 VOLT ERROR: This indicates that the plus 15 volt supply is outside the allowable
10% tolerance levels, which means it is either below +13.5 volts or
above +16.5 volts.
-15 VOLT ERROR: This indicates that the minus 15 volt supply is outside the
allowable 10% tolerance levels, which means it is either above
-13.5 volts or below -16.5 volts.
+5 VOLT ERROR: This indicates that the plus 5 volt supply is outside the allowable
10% tolerance levels, which means it is either below +4.5 volts or
above +5.5 volts.
SENS. HEAD:
This status field monitors the status of the MPR E-Scan unit sensing head voltage levels. The
following errors can be displayed in this status field:
MEAS. VOLT ERROR: This indicates that the measurement voltage in the sensing
head is out of specification and is either below 0.5 volts or above
9.5 volts.
TEMP. VOLT ERROR: This indicates that the temperature reading derived from the
sensing head voltage is out of specification and exceeds the
temperature high or low limits set in the Monitor Configuration
Limits Menu.
LAMP. VOLT ERROR: This indicates that the voltage needed to maintain a signal of
proper amplitude for the light source is out of specification.
Below 2.5 or above 4.2 volts D.C.
COMM:
This status field is used to indicate an RS-232 communication error between the Monitor and the
MPR E-Scan units. The following error can be displayed in the status field:
COMM. FAILED: This indicates a communication failure, which could be the result
of a bad serial port, bad wiring, or a dead CPU.
PURGE:
This status field monitors all system steam purge functions. The following errors pertaining to
purging can be displayed in this status field:
PURGE FAILED: This indicates that a proper measurement voltage drop was not
seen by either the MPR E-Scan or the Monitor during the purge
operation and thus intelligent cleaning was not effective.