Service manual

MDS E-SCAN MPR E-SCAN OPERATION/ PURGE SELECTION
27
During the purge process, all
readings and analog output are
“on hold” or frozen.
2.6 Purge Selection
Selecting the “Purge”
option from normal mode
adds purge information to the
current display. When
purging, the ‘H’ system
indicator (measurement hold)
is displayed to the far right of
the displayed reading, and the
measurement voltage as well
as the current purge mode is
displayed at the bottom. The
timing information for
purging is set in the
calibration limits menu.
Each purge operation mode can be aborted before the selected time expiration by
selecting the ESC button from the Front Panel at which time the system will resume
normal operation. Following the termination of the purge cycle, the ‘H’ indicator and the
information at the bottom is removed. With intelligent cleaning activated, successive
purge cycles will be started if the previous purge was unsuccessful. This will continue for
three entire purge cycles, at which time a ‘P’ system indicator is displayed if all purges
were unsuccessful.
The purging consists of three internal modes:
1. Condensate Drain Time: This time allows any condensation to drain before steam
cleaning if steam is being used as the cleaning agent.
2. Prism Clean Time: This is the actual time during which cleaning of the prism occurs.
3. Hold Delay Time: This time allows a delay to occur before resuming normal
operation.
NOTE: Purging can be automatically initiated by setting up a purge cycle time in the
calibration limits.
Purge Mo
de Screen