MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION
25
The “Relays” option activates the following display:
There are eight total relays, two
of which are used as inputs
coming from the Monitor and
two are unused. The other four
relays are activated when either
measurement alarms are
occurring or when purging.
Using the Front Panel buttons,
all relays except the input relays
can be tested by toggling the
output conditions and visually
checked by examining the relay
LEDS on the interface board.
Also, by supplying an external
voltage, the operator can test the
custom input relays.
The “Anlg. Out" diagnostic option activates the following display:
There are two standard outputs
that produce a 0 - 10 volt and 4-
20 ma signal and are used for
process measurement and temp.
They are displayed as Analog 1
Out and Analog 2 Out
respectively.
Using the Front Panel buttons,
the operator can test the outputs
in min.., mid., and max., value
conditions. Also by editing the
min. and max. values, the
operator can even calibrate the
output to work more accurately
with various analog loads.
Diagnostics Analog Output Screen
Diagnostics Relay Screen