Service manual
MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION
24
2.5 Diagnostic Selection
The Diagnostics menu contains the following options:
1. Voltages - This displays all
pertinent system voltages such as:
measurement voltage, temperature
voltage, and lamp voltage.
2. Relays - This allows the
operator to toggle system relays
for testing or troubleshooting
purposes. The toggled relays can
be visually checked by examining
the relay LEDS on the interface
board. This option also allows the
operator to test custom inputs.
3. Anlg. Out - This option enables the operator to test system output as well as calibrate the
output to work with various analog loads.
4. Time/Date - This option enables the operator to reset the system clock with a valid time and
date. Such adjustments may be necessary as a result of time zone differences or when the system
clock batteries fail.
The “Voltages option results in the following display:
Measurement Voltage: The
voltage produced when measuring a
solution sample. The first voltage
displayed is the raw sensing head
voltage. The voltage in brackets is
the compensated voltage. Range is
(0-10 volts). See also Analog O
menu.
Temperature Voltage: A voltage
that produces the process
temperature readout as follows:
(Voltage - .5v) * 50 = Deg. C.
Lamp Voltage: A nominal voltage
needed to maintain a signal of
proper amplitude.
Diagnostics Mode Screen
Diagnostic Voltages Screen