MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION
23
Note: If the process measurement changes after the sampled reading is recorded then the final
adjusted reading will not reflect the lab reading. However, since the adjustment is calculated
from the recorded sampled reading, the final reading will be accurate and will not require further
manual adjustment. Thus the lab reading should only be thought of as the desired reading if the
process is stable. In all other cases, the lab reading simply provides a means of calculating an
error adjustment from a previously recorded reference point.
Manual Adjustment
The operator should proceed with caution when adjusting analog zero manually since this will
change the output of the instrument. To reduce the likelihood of manual adjustment error, it is
suggested that the default least significant digit be altered first so the operator can observe the
effects on the measurement reading. Pressing the up or down arrow buttons will output the
altered analog zero adjustment.
2.4 Print Selection
Prints all System Calibration and Configuration Settings.