Service manual
MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION
19
Measurement Alarm Divert: If the measurement falls below this limit, a system alarm is
activated. An error message will appear in the status window in normal mode, and the divert
alarm relay will be turned off to activate a divert to the Monitor.
Measurement Alarm Low: If the measurement drops below this limit, a system alarm is
activated. An error message will appear in the status window in normal mode, and the low
measurement alarm relay will be turned on.
Measurement Alarm Delay: This is the alarm delay time in seconds for the process
measurement that must expire while a limit is exceeded before an error is activated.
Purge Cycle Time: This is the time interval in minutes between automatic purging. During
purging, all system readings and analog outputs are frozen. A “0" setting disables automatic
purging.
Condensate Drain Time: When the purge cycle is activated, this is the time in seconds that
elapse before a prism clean occurs. During this time the condensate drain relay is activated and
condensation is allowed to drain before steam cleaning.
Prism Clean Time: This is the time in seconds during which cleaning of the prism occurs and
the prism clean relay is activated.
Hold Delay Time: This is the time in seconds that elapse after a prism clean occurs. This time
allows a delay to occur before resuming normal process measurement. Readings are held
constant during the period.
Meas. Anlg. Min. Output: This is the measurement reading that will produce a 4 ma or 0 volt
output.
Meas. Anlg. Max. Output: This
is the measurement reading that
will produce a 20 ma or 10 volt
output.
Filter Weight: This is a value
that controls system response
time to process measurement
changes. The larger the value the
slower the response time. The
filter weight divided by 10 will
give approximate response time
in seconds to achieve 63% output
of a step input change.
Calibration Limits Screen 2