Service manual

MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION
18
2.3 Calibration Selection (Note: To access calibration enter password “0000")
The calibration menu contains
four menu options:
1. Limits - This option contains
the system limits for alarms,
purge cycle, analog output and
filter weight.
2. Meas. - This option allows
entry to the calibration
measurement routine. This is used
to calibrate the process
measurement so that a linear
interpolation can be made to get
the Refractive Index based on the
current measurement voltage.
3. Temp. - This option allows
entry to the calibration temperature compensation routines. This is used to compensate the
process measurement changes that are caused by temperature changes.
4. Anlg. 0 (Analog Zero) - This option is used to adjust or “zero in” the instrument to match a
target value.
Using the Front Panel buttons, the operator can highlight and then select the specified calibration
menu options. Selecting the “Limits” menu option would generate the following display:
Using the Front Panel arrow
buttons, the operator can scroll
through and edit selected system
limits. In this example,
Measurement Alarm Divert can
now be changed using the
RIGHT ARROW button.
Calibration Mode Screen
Calibration Limits Screen 1