1.5 Cooling System Requirements
When supplied with a vortex cooler a clean and dry air supply of the proper pressure
(approximately 90 PSI at 8 cubic ft/min.) is required. If equipped with the cooling fan option the
console must be mounted in a controlled environment.
1.6 Start-Up
Turn power on. Allow sufficient time for the sensing head to stabilize at the process
temperature. It is recommend that no adjustments be made for at least 15 minutes after start-up.
Compare the reading against a sample taken from the process line close to where the unit is
installed and at the process operating temperature. If the sample does not equal the displayed
reading, adjust the analog zero per Operation/Calibration section.
1.7 Outputs
1.7.1 Relays
All alarm contacts are 5A, 250 VAC, resistive.
1.7.2 Printer Output
The MDS E-Scan includes a printer mounted in the cabinet.
1.7.3 Analog Outputs
The MDS E-Scan offers current or voltage outputs for the measurement. The standard output for
these readings is a 4 to 20 ma signal that is referenced to ground (non-isolated) so that any device
that is connected to these outputs must have a floating (or isolated) input. Isolated current output
modules are available as an option.
Important: Prior to shipment the sensing head is matched to the console and factory
calibrated. DO NOT ATTEMPT TO RE-CALIBRATE