SECTION 6: Maintenance and Troubleshooting
Preventive Maintenance
WARNING LMA requires periodic inspection and maintenance to perform optimally and achieve maximum
operation life.
WARNING The intervals shown are recommended. Maintenance schedules should be more frequent for
higher use areas.
Area Maintenance
Period
(Month)
1
VESA Adapter
Check if the adapter plate and the monitor are securely attached. Ensure the four screws on the
mounting plate(s) are fastened to the tightest possible position.
Visually inspect for any signs of grinding and gapping.
4
2
MRS Rail Mount Adapter
Ensure the two set screws on the channel are fastened to the tightest possible position.
1
3
Swivel Joint
Ensure screw on bottom of joint is fastened to the tightest possible position using a #3 Philips head
screwdriver.
Ensure set screw at back of joint is tightened. Use a 2.5 mm HEX key to tighten as needed.
1
Troubleshooting
Symptom Possible cause Solution
Too easy or too dicult to tilt the monitor
up and down
Excessively loosened or tightened
hardware
Adjust respective screws to control the
correct movement
Too easy or too dicult to utilize the
portrait/landscape feature
Weight of the monitor is not compatible
with the arm
Please contact Amico Accessories Inc.
The entire arm is rocking slightly when in
use
Damaged hardware Please contact Amico Accessories Inc.
Length of the extension arm is at an angle
Loosened/damaged hardware
Perform preventive maintenance
Check all screws are secured
Please contact Amico Accessories Inc.
Extension arm is not secure or it is rocking
slightly
Arm is too dicult to maneuver with the
monitor
Damaged extension arm
Weight of the monitor exceeds the load
rating
Please contact Amico Accessories Inc.
Arm falls to the lowest position on its own
when monitor is attached
Counterbalance is not adjusted properly
for devices
Adjust for the correct counterbalance
See section 4: adjustment procedure
Excessive gapping or angling at joints of
arm
Damaged hardware or other physical
damage
Equipment is too heavy
Check device mounted for weight
compatibility. If problem persists
contact Amico Accessories Inc.
If the above solutions do not solve your symptoms or you are in need of parts/hardware, please contact Amico
Accessories, 1.877.264.2697