Operating & Maintenance Manual Gas Control Panel w w w . a m i c o .
Contents 2 Intended Use/Introduction 3 Operation 4 Installation 4 Parts List 4-5 Notes 6-7 Amico Pipeline
Intended Use The Amico Gas Control Panel is designed to deliver Nitrogen or Instrument Air to turbo-surgical tools. Introduction The Gas Control Panel consists of a shut-off valve, delivery and supply pressure gauges, pressure regulator and DISS outlet, in addition an outlet pipe is supplied that can be piped to additional gas outlets. The Gas Control Panel comes in two styles: “Relieving” and “Non-Relieving”.
Operation 1. Turn the Shut-off valve to the “Open” position. 2. Verify that the correct supply pressure is available on the supply gauge. 3. Attach a tool to the DISS outlet and tighten the nut (do not over tighten) or HP “Safety-Swing Coupling” outlet. 4. Adjust the pressure regulator to the required pressure by observing the outlet gauge. 5. Operate the surgical tool and observe the outlet pressure and adjust if needed. 6.
Diagrams 1 1 2 3 10 Gas Outlet 4 13 Adjustable Mounting Bracket 5 11 Disconnect Here 9 12 Gas Inlet 8 7 6 Not visible 15 14 16 Item 1 2 Description 2 1/2" Gauge, 0-400 psi, 1/4" Center Back Mount Brass Tee 1/4" FNPT * 3/8" Tube Part Number Item Description Part Number M-X-MAN-52B 9 Elbow 90° 1/4" MNPT * 1/8" Tube N-X-ELB-ALM2508 N-X-TEE-AL-F304 10 Elbow 90° 1/4" FNPT * 1/8" Tube N-X-ELB-ALF2508 Elbow 90° 1/4" FNPT * 3/8" Tube N-X-ELBOW-AL 3a.
Notes 6 Amico Pipeline
Notes www.amico.
www.amico.com Amico Pipeline | www.amico.com 85 Fulton Way, Richmond Hill Ontario, L4B 2N4, Canada Toll Free Tel: 1.877.462.6426 Toll Free Fax: 1.866.440.4986 Tel: 905.764.0800 Fax: 905.764.0862 Email: info@amico.com APE-OM-GAS-CTRL-PNL 06.20.