Operation, Maintenance & Installation Manual Emergency Oxygen Inlet Station Low Pressure 2” Recessed Mount Surface Mount Model#: M-FILL-OXY-LP2 Model#: M-FILL-OXY-LP2S w w w . a m i c o .
Contents Introduction 3 Medical Facility Responsibility 3 Installation of Emergency Oxygen Inlet Station 4 Operating Instructions 4 Repair and Replacement Parts List 5 Schematic Drawings 2 6-7 Optional Check and Relief Valves Parts List 7 Model Numbers Also Available 7 Amico Pipeline
Introduction As required by NFPA-99 (National Fire Protection Association), Health Care Facilities oxygen systems having the source of supply outside the building shall have an inlet to connect a temporary, auxiliary source of supply for emergency or maintenance situations. This connection shall include necessary valves and to allow for an emergency supply of oxygen and isolation of the normal source of supply.
Installation of Emergency Oxygen Inlet Station NOTE: Prior to installation, inspect the unit to ensure no tampering or damages were incurred. 1. Anchor the oxygen inlet station on a finished wall with the appropriate hardware. 2. Before brazing, wrap wet rags around the tube extension next to the valve body to prevent overheating and possible damage to the valve seals. Connect copper tubing to the copper extension using brazing methods and materials in accordance with NFPA-99 or CSA Z7396.1. 3.
Repair and Replacement Parts List: 1. V-X-RK-NIB-20 Repair kit for 2” ball valve - seats and O-rings 2. V-X-VLV-HAND-L Valve handle sleeve for 2” 3. V-X-GAUGE-100 Gauge 100 psi/690 kPa – 1/8” NPT 4. M-X-OXF-20-PLG 2” NPT hex-head plug 5. V-X-BOX-PLUG-25-B Box plug for oxygen inlet station 2 ½” (black) 6.
Schematic Drawing - Recessed Mount Typical Piping Connect to pipeline 25.5 [648] 8 12" [305] [203] Check Valve Emergency Low Pressure Oxygen Inlet Box. Main Oxygen Supply Shutoff Valve 28 18-1/2" [711] [470] 27 17-1/2" [445] [686] Relief Valve 75 PSIG Turn Down and Screen Pressure Switch Main Line Shutoff Valve Provision for Padlock 16 12" [406] [305] 13-1/2" 17.5 14-1/2" 18.5 [343] [445] [368] [470] Inlet Oxygen connection 2" NPT 7-1/4" 8 [184] [203] 2" Type "K" Copper outlet pipes.
Schematic Drawing - Surface Mount Typical Piping Turn Down and Screen Relief Valve 75 PSIG Emergency Low Pressure Oxygen Inlet Box Shutoff Valve 26" [660] Shutoff Valve 26 [660] Main Oxygen Supply Pressure Switch Check Valve 8 [203] 8" [203] 11 [279.
www.amico.com Amico Pipeline | www.amico.com 85 Fulton Way, Richmond Hill Ontario, L4B 2N4, Canada Toll Free Tel: 1.877.462.6426 Toll Free Fax: 1.866.440.4986 Tel: 905.764.0800 Fax: 905.764.0862 Email: info@amico.com APE-OM-INSTAL-EOIS-LP-2IN 01.07.